Vikingegaarden A/S

Height above ground and why 

 

The temperature sensor must be placed at a height of 2 meters – this is the official measuring height. It can be placed at a different height, but this is only in connection with special requests. The temperature in the weather forecasts is a 2 meter temperature. Another reason is that the temperature can vary with altitude, especially in winter when it is clear and windless. It can be around 4 degrees colder at ground level than at a height of 2 metres. If the temperature sensor is at a different height, this must be noted. It is possible on the basis of other parameters to determine the temperature, e.g. down at the ground surface. 

 

The stations can be equipped with either one of the brands MetSpec, NGS or Vaisala, which is built together with the humidity sensor – see below. 

 

Sheath/steel protection 

 

Cover in which the sensor is mounted protects against solar radiation, 

and which at the same time ensures good ventilation. 

EE060 from E+E Elektronik 

 

 

 

 

 

EE060 from E+E Elektronik 

 

Specifications 

Temperature 

Operating range -40 to +60 °C 

Accuracy ±0.3 gr. °C – at 20 gr. 

 

Humidity 

±2.5% at 20 °C 

 

 

 

 

 

 

 

HMS83 from Vaisala 

 

Specifications

Temperature 

Operating range -40 to +60 °C 

Accuracy ±0.3 gr. °C – at 20 gr. 

 

Humidity 

At temp. between +10 and +30°C 

 

0-90% ±3% 

90-100% ±5% 

 

At temp. between -20 and +10°C 

0-90% ±5% 

90-100% ±7% 

 

At temp. between -40 and -20 °C 

0-100% ±7% Humidity ±2.5% RH at 20 gr. 

 

 

 

 

 

HMP155 from Vaisala 

The sensor may be supplied with a yellow protective cap – it must be dismantled. 

 

Specifications 

Operating range -39.2 to +60°C 

Accuracy 0.2°C 

 

The current version 

The previous version 

The sensor can be pulled off for replacement or adjustment 

 

 

 

 

 

 

 

Soil temperature 

 

The sensor for measuring the soil temperature must be buried 10 cm – under normal conditions. 

 

Specifications 

Operating range -55 to +125 °C 

Accuracy 0.5 °C 

 

The wire cannot be removed from the sensor 

 

 

 

 

 

NGS sensor 

 

Specifications 

Operation range -40 to 52.5°C 

Accuracy 1.1°C 

Response time approx. 5 minutes 

Wire connection Ready – signal – AI1 

Black – minus – AGND 

Red – plus – DCOUT 

 

  The cable cannot be dismantled on the sensor.